Indian Journal of Respiratory Care

IJRC Email      Register      Login

VOLUME 3 , ISSUE 1 ( January-June, 2014 ) > List of Articles

Original Article

Graded oxygen delivery using low flow rotameter during positive pressure ventilation using self-inflating bag with a leak - an experimental study

Swathy Mohan, Tisha Ann Skariah, Eribaweimon Shilla, Leslie Lewis

Keywords : Oxygen delivery, self-inflating bag, positive pressure ventilation

Citation Information : Mohan S, Skariah TA, Shilla E, Lewis L. Graded oxygen delivery using low flow rotameter during positive pressure ventilation using self-inflating bag with a leak - an experimental study. Indian J Respir Care 2014; 3 (1):370-373.

DOI: 10.5005/ijrc-3-1-370

License: NA

Published Online: 01-12-2022

Copyright Statement:  NA


Abstract

Introduction: Bag-mask ventilation does not always guarantee 100% seal around airway opening and hence delivered fraction of oxygen (FDO2) may vary. Aim: To determine the FDO2 via a standard neonatal selfinflating bag with and without application of leak, without a reservoir bag with flow rates of 0.1, 0.2, 0.4, 0.6, 0.8, 1, 2, 4 and 6 L/min. Method: A test lung with a three way rotator to produce 0%, 30%, 50% and 75% leak was connected to a 3.5 mm ID endotracheal tube (ETT) connected to a VBM standard preterm self-inflating bag of 250 mL without a reservoir. The ETT was connected to ENVENTEC oxygen analyser and the bag inlet to a 100% oxygen source. Ventilation was done to a peak pressure of 15-20 cm H2O, at 40 bpm. FDO2 was recorded every 30 s, until the difference between two consecutive values was ≤1%. Result: The change in FDO2 was maximum when compared between 0.1 L/min and 6 L/min. For all leak percentages, the delivered oxygen showed an increasing trend from 0.1 to 0.6 L/min, a plateau from 0.6 to 1 L/min after which it increased. The change in FDO2 was different for different flow rates among various leak percentages. The highest delivered oxygen without leak was 56.15 ± 3.45% with 4 L/min flow. The oxygen delivered with 70% leak was considerably lower when compared to values with different leaks with respective flow rates. Conclusion: FDO2 increases with oxygen flow rates greater than 1 L/min. Leak around the mask has no effect of FDO2. Equilibration of FDO2 is achieved with in 60-90 s.


HTML PDF Share
  1. Wiswell TE. Neonatal Resuscitation. Respiratory Care 2003; 48(3):288-93.
  2. Johnston KL and Khalid Aziz. The selfinflating resuscitation bag delivers high oxygen concentration when used without a reservoir: Implication for neonatal resuscitation. Respiratory Care 2009; 54:1665-70.
  3. Thió M, Bhatia R, Dawson JA, Davis PG. Oxygen delivery using neonatal self-inflating resuscitation bags without a reservoir. Archives of Disease in Childhood - Fetal and Neonatal Edition 2010 Sep; 95(5):F315-9.
  4. Dawson JA, Davis PG, O’Donnell CPF, Kamlin COF, Morley CJ. Free-flow oxygen delivery to newly born infants. Arch Dis Child Fetal Neonatal Ed 2007; 92:F132–F134.