REVIEW ARTICLE

Blood gas sampling - Pre-analytical issues

Heera Lal Mahto, Sanjay Sasikumar

Keywords : Blood gas sampling, pre-analytical error, pH, PO2, PCO2

Citation Information : Mahto HL, Sasikumar S. Blood gas sampling - Pre-analytical issues. Indian J Respir Care 2017; 6 (1):758-761.

DOI: 10.5005/jp-journals-11010-06103

License: NA

Published Online: 02-12-2022

Copyright Statement:  NA


Abstract

Arterial blood gas analysis is a commonly ordered test in the intensive care unit and the results are required to adjust the ventilator settings through all phases of mechanical ventilation including initiation, maintenance and weaning. Since important clinical decisions are made based on the results of these tests, they have to be very reliable. Many errors can creep up in the process of obtaining the test results. The measurement of the parameters pH, pCO2, and pO2 is vulnerable to a number of pre- analytical errors. Up to 70% of all errors in blood gas testing occurs in the pre-analytical phase and fortunately many of them can be prevented. This article is intended to discuss the pre-analytical issues during blood gas sampling and how to avoid those issues.


HTML PDF Share
  1. Lundberg GD. Acting on significant laboratory results. JAMA 1981;245:1762-3.
  2. Plebani M, Laposata M, Lundberg GD. The brain-to-brain loop concept for laboratory testing 40 years after its introduction. Am J Clin Pathol 2011;136:829-33.
  3. Plebani M, Sciacovelli L, Aita A, Pelloso M, Chiozza ML.. Performance criteria and quality indicators for the pre-analytical phase. Clin Chem Lab Med 2015; 53: 943-48.
  4. Hawkins R. Managing the pre- and postanalytical phases of the total testing process. Ann Lab Med 2012;32:5-16.
  5. Plebani M. The detection and prevention of errors in laboratory medicine. Ann Clin Biochem 2010;47:101-10.
  6. De Koninck AS, De Decker K, Van Bocxlaer J, Meeus P, Van Hoovels L. Analytical performance evaluation of four cartridge-type blood gas analyzers. Clin Chem Lab Med 2012;50:1083-91.
  7. Plebani M. Exploring the iceberg of errors in laboratory medicine. Clin Chim Acta 2009; 404:16-23
  8. Blood gas and pH analysis and related measurements; approved guideline -second edition. Vol. CLSI Document C46A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2010.
  9. Hajiseyedjavady H, Saeedi M, Eslami V, Shahsavarinia K, Farahmand S. Less painful arterial blood gas sampling using jet injection of 2% lidocaine: a randomised controlled clinical trial. Am J Emerg Med 2012; 30: 1100-4.
  10. Toffaletti JG, McDonnell EH. Effect of small air bubbles on changes in pO2 and blood gas parameters: calculated vs. measured effects. www.acutecaretesting.org July 2012.
  11. Wiwanitikit V. Glass syringes are better than plastic for preserving arterial blood gas for oxygen partial pressure determination: an explanation based on nanomaterial composition. Int J Nanomedicine 2006; 1:.223-24.
  12. Higgins C. The use of heparin in preparing samples for blood gas analysis. www. acutecaretesting.org April 2007.
  13. Hutchison AS, Ralston SH, Dryburgh FJ, Small M, and Fogelman I. Too much heparin: possible source of error in blood gas analysis BMJ 1983; 287: 1131-32.
  14. Greenbaum R, Nunn JF, Prys-Roberts C, Kelman GR. Metabolic changes in whole human blood. Respiration Physiology 1967;2: 274-82.
  15. Burnett RW, Covington AK, Fogh-Andersen N et al. International Federation of Clinical Chemistry (IFCC) Approved IFCC recommendations on whole blood sampling, transport and storage for simultaneous determination of pH blood gases and electrolytes. Eur J Clin Cem Clin Biochem 1995: 33: 247-53.
  16. Higgins C. Pneumatic tube transport of blood samples – an update. www.acutecaretesting.org Jan 2005
  17. Lippi G, Fontana R, Avanzani P et al. Influence of spurious hemolysis on blood gas analysis. Clin Chem Lab Med 2013; 51: 1651-4.
  18. Ashwood ER, Kost G, Kenny M. Temperature correction of blood-gas and pH measurements. Clin Chem 1983;29:1877-85.
  19. Verhovsek M, Henderson MP, Cox G, Luo HY, Steinberg MH, Chui DH. Unexpectedly low pulse oximetry measurements associated with variant hemoglobins: A systematic review. Am J Hematol 2010;85:882-5.
  20. Eberhard P, Mindt W. Interference of anesthetic gases at oxygen sensors. Birth Defects Orig Artic Ser 1979;15:65-74.
  21. Scott M, LeGrys V, Klutts J. Electrolytes and blood gases. In: Burtis C, Ashwood E, Bruns D, eds. Tietz textbook of clinical chemistry and molecular diagnostics. St. Louis, MO: Elsevier Saunders, 2006:1008.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.